Technical Specifications
HandyPROBE Technical Specifications
| Weight |
450 g (1 lb) |
| Dimensions |
204 x 159 x 97 mm (8 x 6.26 x 3.8 in) |
| Measurement speed |
30 points/s |
| Operating temperature range |
15-40 °C (59-104 °F) |
| Operating humidity range (non-condensing) |
10-90% |
| Certification |
EN 301 489-1, EN 301 489-3, EN 300 220-1 |
Technical Specifications
|
 |
 |
 |
 |
| ACCURACY (1) |
Up to 0.022 mm (0.0009 in.) |
Up to 0.025 mm (0.0010 in.) |
Up to 0.030 mm (0.0012 in.) |
n/a |
| SINGLE POINT REPEATABILITY (3) |
3.80 m3 (135 ft 3) |
0.045 mm (2) (0.0018 in.) |
0.050 mm (5) (0.0020 in.) |
0.055 mm (5) (0.0022 in.) |
n/a |
7.80 m3 (275 ft3) |
n/a |
0.055 mm (2) (0.0022 in.) |
0.065 mm (5) (0.0026 in.) |
n/a |
14.8 m3 (525 ft3) |
n/a |
n/a |
0.095 mm (2) (0.0037 in.) |
n/a |
| VOLUMETRIC ACCURACY (4) |
3.80 m3 (135 ft3) |
0.075 mm (2) (0.0029 in.) |
0.080 mm (5) (0.0031 in.) |
0.090 mm (5) (0.0035 in.) |
n/a |
7.80 m3 (275 ft3) |
n/a |
0.085 mm (2) (0.0033 in.) |
0.095 mm (5) (0.0038 in.) |
n/a |
14.8 m3 (525 ft3) |
n/a |
n/a |
0.170 mm (2) (0.0067 in.) |
n/a |
| MEASUREMENT SPEED |
30 Hz |
30 Hz |
30 Hz |
30 Hz |
| WEIGHT |
5,5 kg (12 lbs.) |
5,5 kg (12 lbs.) |
5,5 kg (12 lbs.) |
450 g (1 lb.) |
| DIMENSIONS |
1089 x 174 x 119 mm (43 x 6.9 x 4.7 in.) |
1089 x 174 x 119 mm (43 x 6.9 x 4.7 in.) |
1089 x 174 x 119 mm (43 x 6.9 x 4.7 in.) |
204 x 159 x 97 mm (8 x 6.26 x 3.8 in.) |
| OPERATING TEMPERATURE RANGE |
15–40 °C (59-104 oF) |
15–40 °C (59-104 oF) |
15–40 °C (59-104 oF) |
15–40 °C (59-104 oF) |
| OPERATING HUMIDITY RANGE (NON CONDENSING) |
10–90 % |
10–90 % |
10–90 % |
10–90 % |
| CERTIFICATIONS |
EN 301 489-1, EN 301 489-3, EN 300 220-1 |
EN 301 489-1, EN 301 489-3, EN 300 220-1 |
EN 301 489-1, EN 301 489-3, EN 300 220-1 |
EN 301 489-1, EN 301 489-3, EN 300 220-1 |
(1) Volumetric accuracy using dynamic referencing mode and a 500 mm artefact.
(2) Each system tested according test methods given in the ASME B89.4.22 standard.
(3) Single point repeatability: The probe of the HandyPROBE is located within a conical socket. Individual points are measured from multiple approach directions.
(4) Volumetric accuracy: Performance is assessed by measuring traceable length artifacts in different locations and with different orientations within the working volume of the HandyPROBE (maximum range / 2).
(5) Typical value.
Back to Top